SECTION 4 PREPARATION FOR USE AND INSTALLATION
_______________________________________________________________________
4-2
Humidity: The monitor is designed to operate within specifications at a relative non-
condensing humidity of 15% to 95%.
Pressure: The monitor will operate satisfactorily at or above sea level, and is unaffected by
extremes or changes in altitude within atmospheric pressures of 360 mmHg to 800 mmHg.
4.1.3 Power Requirements and System Grounding
The BIS VISTA Monitoring System requires a power source of 100-240 VAC, 50-60Hz.
Current consumption is 0.7 ampere maximum.
To protect operating personnel and patients, the monitor must be properly grounded.
Accordingly, the monitor is equipped with a hospital grade line cord. The power cord
grounds the system to the power line ground when plugged into an appropriate three-wire
receptacle.
WARNING!
FOR PROPER GROUNDING, THE POWER RECEPTACLE
MUST BE A THREE-WIRE GROUNDED OUTLET. A HOSPITAL
GRADE OUTLET IS REQUIRED. NEVER ADAPT THE THREE-
PRONG PLUG FROM THE MONITOR TO FIT A TWO-SLOT
OUTLET. IF THE OUTLET HAS ONLY TWO SLOTS, MAKE
SURE THAT IT IS REPLACED WITH A THREE-SLOT
GROUNDED OUTLET BEFORE ATTEMPTING TO OPERATE
THE MONITOR.
IF THE INTEGRITY OF THE EXTERNAL PROTECTIVE
EARTH GROUND IS IN DOUBT, THE BIS VISTA MONITOR
SHALL BE OPERATED FROM ITS INTERNAL BATTERY
POWER SOURCE ONLY.
FOR BIS VISTA SYSTEMS USED OUTSIDE OF NORTH
AMERICA - A HARMONIZED LINE CORD WITH
CONDUCTORS HAVING A CROSS SECTIONAL AREA
GREATER THAN 0.75 mm
2
MUST BE USED.
4.1.4 Site Preparation: Mounting the Monitor
Aspect Medical Systems, Inc. strongly recommends permanent mounting of the BIS VISTA
monitor to the anesthesia machine to enhance safety and facilitate ease-of-use. Please
contact your local representative or Aspect to discuss mounting options.
WARNING!
BE SURE THE MONITOR IS MOUNTED SECURELY IN PLACE
TO AVOID PERSONAL OR PATIENT INJURY.