SECTION 6 PREVENTIVE MAINTENANCE
_______________________________________________________________________
6-2
6.2.1 Monitor Checkout Procedure
The successful completion of this test will verify that the monitor is functioning and that all
operator input (touch keys) and output (display and audio alarm) are OK.
IF ANY FAILURES ARE NOTED, SEE SECTION 7.3, “BIS
VISTA System Messages And Corrective Actions”.
1. Disconnect the BISx from the monitor.
2. Connect power cord to monitor. Plug power plug into appropriate wall outlet.
Verify that the light to the right of the ON/Standby button is yellow.
3. Start up monitor by pressing the ON/Standby button (lower right corner).
Verify that the light to the right of the ON/Standby button is green.
Verify all self-tests complete successfully. A beep tone sounds.
Verify next screen says “Connect BISx.”
4. Connect BISx to monitor with PIC and sensor.
Verify screen says, “BISx Initialization Complete.”
Verify SENSOR CHECK begins.
5. Disconnect power cord from rear of monitor.
Verify ‘OPERATING ON BATTERY BACKUP’ is displayed.
Verify battery icon displays below BIS number.
NOTE:
Since the BIS VISTA monitor has a built in battery backup, the monitor will power up with
or without AC power applied. This step assures that this part of the test is performed under
battery power
6. Reconnect power cord.
Verify battery icon is not displayed below BIS number.
Verify ‘OPERATING ON BATTERY BACKUP is not displayed.
NOTE:
Since the BIS VISTA monitor has a built-in battery backup, the monitor will power up with
no AC power applied. This step assures that this part of the test is performed in AC
operation.
IF ANY FAILURES ARE NOTED:
Verify AC power outlet (wall outlet) is supplying AC Volts of 110vac to 240vac at a
frequency of 50hz to 60hz. Move power cord to known good outlet.
Verify AC power cord is good. Swap power cord with known good one.