SECTION 6 PREVENTIVE MAINTENANCE
_______________________________________________________________________
6-7
user of the equipment. However, as stated in the operating manual, an enclosure leakage
current test should be performed on the exposed metal parts and should be checked
periodically to ensure that the integrity of the equipment’s insulation system is maintained.
The leakage current test should include measurement of ground wire leakage, enclosure
leakage, and patient leakage.
Ground wire leakage typically can be performed automatically by connecting the A/C power
cord of the BIS VISTA Monitor into a safety tester. The enclosure leakage may be measured
by any safety test equipment that is capable of connecting to isolated conductive parts and
measuring the current from those parts to earth. The patient connection terminals of many
safety testers can be used for this purpose. The patient leakage current of the BISx can be
measured by connecting the patient connection terminals of a safety tester to a Sensor
Simulator that is connected to the PIC.
To check BISx patient isolation:
1. Connect the Sensor Simulator to the PIC in place of a sensor. (See Section 10.2
“Sensor Simulator P/N 186-0137”.)
2. Short the two circular terminals at the end of the simulator using conventional
methods such as jumpers or alligator clips. Wire attached with screws will work also.
3. Connect the test lead to the shorted terminals. Make sure that you are not touching
the simulator beyond this point.
4. Proceed to test instrument for leakage current as per established facility protocols
and procedures for safety testing of medical devices.